• Title of article

    Large volume room temperature gamma-ray spectrometers from CdxZn1−xTe

  • Author/Authors

    Lund، نويسنده , , J.C and VanScyoc III، نويسنده , , J.M. and James، نويسنده , , R.B and McGregor، نويسنده , , D.S and Olsen، نويسنده , , R.W، نويسنده ,

  • Pages
    6
  • From page
    256
  • To page
    261
  • Abstract
    Recent developments in Cd1−xZnxTe (CZT) crystal growth technology have enabled the production of large samples of detector grade material. In this paper we discuss the technical issues associated with building large volume CZT detectors (> 10 cm3) for use in gamma-ray spectroscopy. In particular, an analysis of the electronic noise in CZT devices and how it scales with detector size is presented, along with a discussion of methods to reduce noise in large CZT detector systems. The degradation of energy resolution caused by charge collection problems is also discussed and the use of electronic signal processing and improved device design to minimize charge collection effects are described.
  • Journal title
    Astroparticle Physics
  • Record number