Barber، نويسنده , , H.B and Marks، نويسنده , , D.G and Apotovsky، نويسنده , , B.A and Augustine، نويسنده , , F.L and Barrett، نويسنده , , H.H and Butler، نويسنده , , J.F and Dereniak، نويسنده , , E.L and Doty، نويسنده , , F.P and Eskin، نويسنده , , J.D. and Hamilton، نويسنده , , W.J and Matherson، نويسنده , , K.J and Venzon، نويسنده , , J.E and Woolfenden، نويسنده , , J.M. and Young، نويسنده , , E.T، نويسنده ,
Pages :
4
From page :
262
To page :
265
Abstract :
We report on the construction and initial testing of a 48 × 48 CdZnTe array with 125 μm pixel spacing and multiplexer readout. Large portions of the array function well but there was a loss of pixels near one corner of the array due to non-interconnecting indium bumps. This problem is readily correctable. Excellent single-pixel spectra were obtained with a 99mTc source using an adjustment technique that accounts for energy deposited in neighboring pixels. A point-spread function (PSF) taken at 140 keV yielded a spatial resolution of 230 μm, much better than required for nuclear medicine applications. No problems were found that are not readily correctable or of much less significance for CdZnTe arrays having larger pixel spacing. We are now constructing 64 × 64 CdZnTe arrays with 380 μm pixel spacing for use in an ultra-high resolution brain SPECT imaging system.