• Title of article

    Mercuric iodide photodetector arrays for gamma-ray imaging

  • Author/Authors

    Patt، نويسنده , , B.E and Iwanczyk، نويسنده , , J.S. and Wang، نويسنده , , Y.J and Tornai، نويسنده , , M.P and Levin، نويسنده , , C.S and Hoffman، نويسنده , , E.J، نويسنده ,

  • Pages
    6
  • From page
    295
  • To page
    300
  • Abstract
    We have developed a novel detector device, namely: a mercuric iodide photodetector array (HgI2 PDA) to be used with a scintillating crystal for X-ray and gamma-ray imaging. A HgI2 PDA prototype with 16-pixels has been evaluated with segmented scintillators (CsI(Tl)) coupled to the HgI2 PDA. Pixel energy resolution of 9.4% FWHM has been obtained at 122 keV for a 4 mm thick scintillator corresponding to 72% stopping at 140 keV, and pixel energy resolution of 11% FWHM has been obtained at 122 keV with the 10 mm thick scintillator.
  • Journal title
    Astroparticle Physics
  • Record number

    1999288