Title of article :
Properties of CF4 and isobutane for use in microstrip gas counters
Author/Authors :
Prendergast، نويسنده , , E.P. and Agterhuis، نويسنده , , E.H. and Kuijer، نويسنده , , P.G. and van den Brink، نويسنده , , Jan A. M. de Haas، نويسنده , , A.P. and Kamermans، نويسنده , , R. and de Laat، نويسنده , , C.T.A.M. and Lourens، نويسنده , , W. and Oskamp، نويسنده , , C.J. and Snellings، نويسنده , , R.J.M. and Voerman، نويسنده , , J.M.، نويسنده ,
Pages :
5
From page :
243
To page :
247
Abstract :
The gain of microstrip gas counters (MSGCs) was studied for two of the most commonly used gases, CF4 and isobutane, in the range of 100 to 1015 mbar gas pressure and of 100 to 400 V bias voltage. It was found that the microstrip gain decreases with increasing gas pressure, however the signal induced by non-relativistic ions exhibits a maximum at a gas-dependent pressure. This behaviour is described by computer simulations of the evolution of single-electron avalanches.
Keywords :
GAIN , CF4 , Low-pressure , simulations , MSGC , Isobutane
Journal title :
Astroparticle Physics
Record number :
2000382
Link To Document :