• Title of article

    On the influence of afterpulsing in PMTs on the trigger threshold of multichannel light detectors in self-trigger mode

  • Author/Authors

    Mirzoyan، نويسنده , , R. and Lorenz، نويسنده , , E. and Petry، نويسنده , , D. and Prosch، نويسنده , , C.، نويسنده ,

  • Pages
    5
  • From page
    74
  • To page
    78
  • Abstract
    In some experiments photomultiplier based multichannel detectors are used to measure short light flashes in self-trigger mode in the presence of a significant background illumination. This background can induce a sizeable afterpulse rate. Contrary to published data the afterpulses can have rather large amplitudes and might be interpreted as genuine signals. This is the case with current imaging air Cherenkov telescopes in cosmic ray experiments. We found that due to the simple trigger logic of contemporary telescopes it is not the light of the night sky (LONS), but the afterpulsing which sets the trigger threshold. Afterpulsing features for 5 types of photomultiplier tubes are presented.
  • Keywords
    Trigger threshold , Photomultiplier tubes , Air Cherenkov gamma ray telescopes , High energy astroparticle physics detector , Afterpulsing
  • Journal title
    Astroparticle Physics