Title of article
Monte-Carlo calculation of the spatial dependence of the coaxial HPGe detector efficiency for point sources
Author/Authors
Korun، نويسنده , , M. and Likar، نويسنده , , A. and Vidmar، نويسنده , , T.، نويسنده ,
Pages
6
From page
203
To page
208
Abstract
It is shown that fine geometric modelling of the HPGe detector structure using the GEANT system leads to a good match between the calculated and measured spatial dependence of efficiency for point sources in the region of photon energies from 60 keV to 1115 keV.
Keywords
?-Ray spectroscopy , HPGe detectors , Monte-Carlo calculation
Journal title
Astroparticle Physics
Record number
2001326