Title of article :
Evaluation of a digital X-ray detector based on a phosphor-coated CCD for mammography
Author/Authors :
Taibi، نويسنده , , A. and Del Guerra، نويسنده , , A. and Gambaccini، نويسنده , , M. and Marziani، نويسنده , , M. and Tuffanelli، نويسنده , , A.، نويسنده ,
Pages :
4
From page :
210
To page :
213
Abstract :
The imaging performance of a phosphor-coated CCD system for mammography application is evaluated. The coated CCD is obtained by direct deposition of a Gd2O2S:Eu powder onto a detector surface. Modulation transfer function, noise power spectrum, and detective quantum efficiency are measured. Their dependence on the input beam X-ray energy is discussed. Images of a test phantom for mammography are compared with those obtained with a screen-film combination.
Keywords :
X-ray detector , mammography , CCD
Journal title :
Astroparticle Physics
Record number :
2001702