Title of article
Time resolution and linearity measurements for a scintillating fiber detector instrumented with VLPCs
Author/Authors
Bross، نويسنده , , A and Chaves، نويسنده , , A and Costa، نويسنده , , J and Johnson، نويسنده , , M and Moreira، نويسنده , , L and Thompson، نويسنده , , J، نويسنده ,
Pages
10
From page
87
To page
96
Abstract
The time resolution for charged particle detection is reported for a typical scintillating fiber detector instrumented with Rockwell HISTE-IV Visible Light Photon Counters (VLPC). The resolution measurements are shown to agree with a simple Monte-Carlo model, and the model is used to make recommendations for improved performance. In addition, the gain linearity of a sample of VLPC devices was measured. The gain is shown to be linear for incident light intensities which produce up to approximately 600 photoelectrons per event.
Journal title
Astroparticle Physics
Record number
2001929