    Time resolution and linearity measurements for a scintillating fiber detector instrumented with VLPCs

  • Author/Authors

    Bross، نويسنده , , A and Chaves، نويسنده , , A and Costa، نويسنده , , J and Johnson، نويسنده , , M and Moreira، نويسنده , , L and Thompson، نويسنده , , J، نويسنده ,

  • Pages
    10
  • From page
    87
  • To page
    96
  • Abstract
    The time resolution for charged particle detection is reported for a typical scintillating fiber detector instrumented with Rockwell HISTE-IV Visible Light Photon Counters (VLPC). The resolution measurements are shown to agree with a simple Monte-Carlo model, and the model is used to make recommendations for improved performance. In addition, the gain linearity of a sample of VLPC devices was measured. The gain is shown to be linear for incident light intensities which produce up to approximately 600 photoelectrons per event.