• Title of article

    Efficiency calibration of a Si(Li) detector in the low-energy region using electron bremsstrahlung

  • Author/Authors

    Tschischgale، نويسنده , , J and Küchler، نويسنده , , D and Lehnert، نويسنده , , U and Zschornack، نويسنده , , G، نويسنده ,

  • Pages
    5
  • From page
    387
  • To page
    391
  • Abstract
    In the low-energy region the determination of the detector efficiency is complicated by the fact of nonavailability of certificated calibration sources. To solve this problem the use of electron bremsstrahlung for determining the relative photon detection efficiency of a Si(Li) detector in the low-energy range 0.5–19 keV is proposed. The calibration is based on energy dispersive measurements of electron bremsstrahlung emitted by an X-ray tube. Model spectra are computed from thick-target bremsstrahlung spectra in a semiclassical approximation and from a physical model for the detector response. Model spectra are fitted to the measured spectra using a parametric adaptor term. Thereby, the error of the determined detector efficiency is approximately 5%.
  • Journal title
    Astroparticle Physics
  • Record number

    2003188