Author/Authors :
Morhac، نويسنده , , Miroslav and Kliman، نويسنده , , J?n and Matou?ek، نويسنده , , Vladislav and Veselsk?، نويسنده , , Martin and Turzo، نويسنده , , Ivan، نويسنده ,
Abstract :
The use of deconvolution operation makes it possible to improve the resolution in γ-ray spectra. The paper describes a non-oscillating method of Gold deconvolution. The method is generalized for multidimensional spectra. A new optimized algorithm of one- and two-dimensional Gold deconvolution aimed to reduce number of computer operations has been derived in the paper.