• Title of article

    Polarimetry on dense samples of spin-polarized 3He by magnetostatic detection

  • Author/Authors

    Wilms، نويسنده , , E and Ebert، نويسنده , , M and Heil، نويسنده , , R. Surkau، نويسنده , , R، نويسنده ,

  • Pages
    8
  • From page
    491
  • To page
    498
  • Abstract
    A very sensitive low-field fluxgate magnetometer is used to detect the static magnetic field produced by dense samples of spin-polarized 3He gas contained in spherical glass cells at pressures around several bars. The 3He nuclear polarization can be extracted with high precision ΔPP < 1% by utilizing magnetostatic detection in combination with adiabatic fast-passage spin reversal. The polarization losses can be kept well below 0.1% thus making this type of polarimetry almost non-destructive. More simply even, P can be measured with reduced accuracy by the change of field when the cell is removed from the fluxgate. In this case the accuracy is limited to about 10% due to the uncertainties about the susceptibilities of the cell walls.
  • Journal title
    Astroparticle Physics
  • Record number

    2003411