Title of article :
In application specific integrated circuit and data acquisition system for digital X-ray imaging
Author/Authors :
D. and Beuville، نويسنده , , E and Cederstrِm، نويسنده , , B and Danielsson، نويسنده , , M and Luo، نويسنده , , L and Nygren، نويسنده , , D and Oltman، نويسنده , , E and Vestlund، نويسنده , , J، نويسنده ,
Pages :
6
From page :
337
To page :
342
Abstract :
We have developed an Application Specific Integrated Circuit (ASIC) and data acquisition system for digital X-ray imaging. The chip consists of 16 parallel channels, each containing preamplifier, shaper, comparator and a 16 bit counter. We have demonstrated noiseless single-photon counting over a threshold of 7.2 keV using Silicon detectors and are presently capable of maximum counting rates of 2 MHz per channel. The ASIC is controlled by a personal computer through a commercial PCI card, which is also used for data acquisition. The content of the 16 bit counters are loaded into a shift register and transferred to the PC at any time at a rate of 20 MHz. The system is non-complicated, low cost and high performance and is optimised for digital X-ray imaging applications.
Journal title :
Astroparticle Physics
Record number :
2004227
Link To Document :