• Title of article

    The study of α/γ ratio for inorganic scintillation detectors

  • Author/Authors

    Sysoeva، نويسنده , , E.V. and Tarasov، نويسنده , , V.A and Zelenskaya، نويسنده , , O.V and Sulyga، نويسنده , , V.A، نويسنده ,

  • Pages
    5
  • From page
    274
  • To page
    278
  • Abstract
    To optimize characteristics of α-radiation detectors as well as those intended to stabilize γ spectrometers, the α/γ ratio has been studied for inorganic scintillation detectors within a wide temperature range (from −20 to +80°C) and at various signal formation time values (from 0.2 to 20 μs). Inorganic scintillators with high atomic numbers have been studied, namely, alkali halide and oxide crystals as well as a semiconducting single-crystal ZnSe(Te). Though the scintillators on the base of the oxide materials have satisfactory resolution and light output, they cannot be considered as efficient detectors of α-radiation due to a low α/γ ratio (∼0.20). The most appropriate detector of α-radiation remains single crystalline CsI(Tl) having the best scintillation characteristics, nonhygroscopic, convenient to use and comparatively cheap.
  • Keywords
    ?/? Ratio , Energy resolution , Signal formation time , Light output , scintillation detector , temperature dependence
  • Journal title
    Astroparticle Physics
  • Record number

    2006255