• Title of article

    Autoradiographic film evaluated with desktop scanner

  • Author/Authors

    Nageldinger، نويسنده , , Guido and Flowers، نويسنده , , Alan and Schwerdt، نويسنده , , Christian and Kelz، نويسنده , , Romano، نويسنده ,

  • Pages
    9
  • From page
    516
  • To page
    524
  • Abstract
    A procedure is described herein for the densitometric calibration of autoradiographic films, utilizing a desktop scanner. The method yields a calibration accuracy of ±0.03 reflection density units over the range from 0 to 1.7. The calibration procedure was applied to Amersham “Hyperfilm-βmax” autoradiographic film against 14C, 36Cl, 60Co, 133Ba, 137Cs and 241Am. An exponential function was used to describe the useful range of the characteristic curves of this film, which is between 1×104 and 1×105 (discharges mm−2) for 14C and 36Cl and between 105–106 (discharge mm−2) for the other gamma-emitting isotopes.
  • Keywords
    Quantitative autoradiography , Desktop scanner , Grey scale , Amersham hyperfilm-?max
  • Journal title
    Astroparticle Physics
  • Record number

    2006752