• Title of article

    Mixed-field dosimetry measurement of a target assembly for an accelerator-based neutron source for boron neutron capture therapy

  • Author/Authors

    Reed، نويسنده , , M.K. and Dobelbower، نويسنده , , M.C. and Woollard، نويسنده , , J.E. and Blue، نويسنده , , T.E، نويسنده ,

  • Pages
    7
  • From page
    160
  • To page
    166
  • Abstract
    The objective of this work was to measure the neutron and gamma-ray absorbed dose components for a target assembly for an Accelerator-Based Neutron Source (ABNS) for Boron Neutron Capture Therapy (BNCT), and to compare these measurements with MCNP calculations in order to verify the calculations of the in-air neutron and gamma-ray absorbed dose components. The measurements were made using the paired ionization chamber technique. The measured neutron and gamma-ray specific absorbed dose components agreed with calculations within experimental errors, which were approximately 10%. The measured gamma-ray specific absorbed dose rate of 140 cGy s−1 A−1 ±12% is consistent with reported yields of 478 keV gamma rays due to the 7Li(p, p′)7Li* reaction. This specific gamma absorbed dose rate is significant and should not be neglected in moderator assembly design.
  • Keywords
    BNCT , Mixed field dosimetry
  • Journal title
    Astroparticle Physics
  • Record number

    2007246