Title of article :
Mixed-field dosimetry measurement of a target assembly for an accelerator-based neutron source for boron neutron capture therapy
Author/Authors :
Reed، نويسنده , , M.K. and Dobelbower، نويسنده , , M.C. and Woollard، نويسنده , , J.E. and Blue، نويسنده , , T.E، نويسنده ,
Pages :
7
From page :
160
To page :
166
Abstract :
The objective of this work was to measure the neutron and gamma-ray absorbed dose components for a target assembly for an Accelerator-Based Neutron Source (ABNS) for Boron Neutron Capture Therapy (BNCT), and to compare these measurements with MCNP calculations in order to verify the calculations of the in-air neutron and gamma-ray absorbed dose components. The measurements were made using the paired ionization chamber technique. The measured neutron and gamma-ray specific absorbed dose components agreed with calculations within experimental errors, which were approximately 10%. The measured gamma-ray specific absorbed dose rate of 140 cGy s−1 A−1 ±12% is consistent with reported yields of 478 keV gamma rays due to the 7Li(p, p′)7Li* reaction. This specific gamma absorbed dose rate is significant and should not be neglected in moderator assembly design.
Keywords :
BNCT , Mixed field dosimetry
Journal title :
Astroparticle Physics
Record number :
2007246
Link To Document :
بازگشت