Title of article
Accuracy and precision of loss-free counting in γ-ray spectrometry
Author/Authors
Pommé، نويسنده , , S. and Alzetta، نويسنده , , J-P. and Uyttenhove، نويسنده , , J. and Denecke، نويسنده , , B. and Arana، نويسنده , , G. and Robouch، نويسنده , , P.، نويسنده ,
Pages
7
From page
388
To page
394
Abstract
The performance of a “Loss-Free Counting” (LFC) method for pulse loss compensation is tested on two HPGe detectors; one with classical RC feedback preamplifier and another with Transistor Reset Preamplifier (TRP). Results are excellent on both, though extra fine-tuning precautions are required for the latter. Pulse loss by pileup is at the origin of an increased count variance in LFC spectra. A new formula for LFC uncertainty is presented and its validity demonstrated for a HPGe detector set-up.
Keywords
nuclear , ?-ray spectrometry , statistics , uncertainty , quality control , Pileup , Loss-free counting
Journal title
Astroparticle Physics
Record number
2008175
Link To Document