Title of article
An innovative method for extracting isotopic information from low-resolution gamma spectra
Author/Authors
Miko، نويسنده , , David and Estep، نويسنده , , Robert J and Rawool-Sullivan، نويسنده , , Mohini W، نويسنده ,
Pages
5
From page
433
To page
437
Abstract
A method is described for the extraction of isotopic information from attenuated gamma-ray spectra using the gross-count material basis set (GC-MBS) model. This method solves for the isotopic composition of an unknown mixture of isotopes attenuated through an absorber of unknown material. For binary isotopic combinations the problem is nonlinear in only one variable and is easily solved using standard line optimization techniques. Results are presented for NaI spectrum analyses of various binary combinations of enriched uranium, depleted uranium, low burnup Pu, 137Cs, and 133Ba attenuated through a suite of absorbers ranging in Z from polyethylene through lead. The GC-MBS method results are compared to those computed using ordinary response function fitting and with a simple net peak area method. The GC-MBS method was found to be significantly more accurate than the other methods over the range of absorbers and isotopic blends studied.
Keywords
Gross-count material basis set (GC-MBS) , Gamma-ray spectra , isotopes
Journal title
Astroparticle Physics
Record number
2008190
Link To Document