• Title of article

    A new method for determination of gamma-ray spectrometer non-linearity

  • Author/Authors

    Fazekas، نويسنده , , B. and Révay، نويسنده , , Zs. and ضstِr، نويسنده , , J. and Belgya، نويسنده , , T. and Molnلr، نويسنده , , G. and Simonits، نويسنده , , A.، نويسنده ,

  • Pages
    5
  • From page
    469
  • To page
    473
  • Abstract
    A new method was introduced to determine the interpolation curve for system non linearity of gamma-ray spectrometers from independent multi-gamma source measurements. The mathematical description of the method and its implementation in the Hypermet-PC program are presented. The performance is illustrated in two different energy ranges, used in NAA and PGAA respectively. As a test case, the energies of 192Ir lines have been determined and compared with the literature values.
  • Keywords
    System non-linearity , Gamma-ray spectrometers
  • Journal title
    Astroparticle Physics
  • Record number

    2008202