• Title of article

    Charge-trap correction and radiation damage in orthogonal-strip planar germanium detectors

  • Author/Authors

    Hull، نويسنده , , E.L. and Jackson، نويسنده , , E.G. and Lister، نويسنده , , C.J. and Pehl، نويسنده , , R.H.، نويسنده ,

  • Pages
    5
  • From page
    125
  • To page
    129
  • Abstract
    A charge-carrier trap correction technique was developed for orthogonal strip planar germanium gamma-ray detectors. The trap corrector significantly improves the gamma-ray energy resolution of detectors with charge-carrier trapping from crystal-growth defects and radiation damage. Two orthogonal-strip planar germanium detectors were radiation damaged with 2-MeV neutron fluences of ~8×109 n/cm2. The radiation-damaged detectors were studied in the 60–80 K temperature range.
  • Keywords
    Charge-carrier trapping , Germanium detector , HPGe , Radiation damage
  • Journal title
    Astroparticle Physics
  • Record number

    2008732