Title of article
Behavior of a trapezoid-based data acquisition system up to 100 kHz and beyond
Author/Authors
Schmidt، نويسنده , , S. and Heftrich، نويسنده , , T. and Glorius، نويسنده , , J. and Hampel، نويسنده , , G. and Plag، نويسنده , , R. and Reifarth، نويسنده , , R. and Slavkovskل، نويسنده , , Z. and Sonnabend، نويسنده , , K. and Stieghorst، نويسنده , , C. and Wiehl، نويسنده , , N. and Zauner، نويسنده , , S.، نويسنده ,
Pages
5
From page
192
To page
196
Abstract
In this work, we investigated the ability of a high-purity germanium detector connected to a trapezoid-filter-based data acquisition system to reliably record signals in spite of high sample activities. By activating multiple Na2CO3 samples with different Na content, we were able to deduce efficiency, resolution and dead time of the system used as a function of the sample activity. Based on the results, we were able to find a setting which allows measurements of event rates up to 35 kHz per readout channel with an energy resolution of 0.3% at the 2754 keV 24Na line.
Keywords
Activation method , Digital data acquisition , Trapezoid filter
Journal title
Astroparticle Physics
Record number
2009641