Title of article :
Solid-state detector system for measuring concentrations of tritiated water vapour and other radioactive gases
Author/Authors :
Nunes، نويسنده , , J.C. and Surette، نويسنده , , R.A. and Wood، نويسنده , , M.J.، نويسنده ,
Pages :
22
From page :
505
To page :
526
Abstract :
A detector system was built using a silicon photodiode plus preamplifier and a cesium iodide scintillator plus preamplifier that were commercially available. The potential of the system for measuring concentrations of tritiated water vapour in the presence of other radioactive sources was investigated. For purposes of radiation protection, the sensitivity of the detector system was considered too low for measuring tritiated water vapour concentrations in workplaces such as nuclear power plants. Nevertheless, the spectrometry capability of the system was used successfully to differentiate amongst some radioactive gases in laboratory tests. Although this relatively small system can measure radioactive noble gases as well as tritiated water vapour concentrations, its response to photons remains an issue.
Keywords :
Solid-state detector , photodiode , HTO-in-air , Airborne radioactivity