Title of article :
Measurements on a prototype segmented Clover detector
Author/Authors :
Shepherd، نويسنده , , S.L and Nolan، نويسنده , , P.J and Cullen، نويسنده , , D.M and Appelbe، نويسنده , , D.E and Simpson، نويسنده , , J and Gerl، نويسنده , , J and Kaspar، نويسنده , , M and Kleinboehl، نويسنده , , A and Peter، نويسنده , , I and Rejmund، نويسنده , , M and Schaffner، نويسنده , , H and Schlegel، نويسنده , , C and de France، نويسنده , , G، نويسنده ,
Pages :
14
From page :
373
To page :
386
Abstract :
The performance of a segmented Clover germanium detector has been measured. The segmented Clover detector is a composite germanium detector, consisting of four individual germanium crystals in the configuration of a four-leaf Clover, housed in a single cryostat. Each crystal is electrically segmented on its outer surface into four quadrants, with separate energy read-outs from nine crystal zones. Signals are also taken from the inner contact of each crystal. This effectively produces a detector with 16 active elements. One of the purposes of this segmentation is to improve the overall spectral resolution when detecting γ radiation emitted following a nuclear reaction, by minimising Doppler broadening caused by the opening angle subtended by each detector element. Results of the tests with sources and in beam will be presented. The improved granularity of the detector also leads to an improved isolated hit probability compared with an unsegmented Clover detector.
Keywords :
Segmented Clover germanium detector , Doppler broadening , Isolated hit probability
Journal title :
Astroparticle Physics
Record number :
2010009
Link To Document :
بازگشت