Title of article
Model-based pulse shape correction for CdTe detectors
Author/Authors
Bargholtz، نويسنده , , Chr and Fumero، نويسنده , , E and Mهrtensson، نويسنده , , L، نويسنده ,
Pages
13
From page
399
To page
411
Abstract
We present a systematic method to improve energy resolution of CdTe-detector systems with full control of the efficiency. Sampled pulses and multiple amplifier data are fitted by a model of the pulse shape including the deposited energy and the interaction point within the detector as parameters. We show the decisive improvements of spectral resolution and photo-peak efficiency that is obtained without distortion of spectral shape. The information concerning the interaction depth of individual events can be used to discriminate between beta particles and gamma quanta.
Keywords
Gamma-ray spectrometry , Pulse-shape analysis , CdTe detector
Journal title
Astroparticle Physics
Record number
2010013
Link To Document