Title of article :
High-quality polycristalline CVD diamond dosimeter: Low bias operation with intensity modulated photons and 62 MeV protons
Author/Authors :
Zani، نويسنده , , Margherita and Bucciolini، نويسنده , , Marta and De Sio، نويسنده , , Antonio and Mori، نويسنده , , Riccardo and Pace، نويسنده , , Emanuele and Scaringella، نويسنده , , Monica and Talamonti، نويسنده , , Cinzia and Tozzetti، نويسنده , , Lorenzo and Bruzzi، نويسنده , , Mara، نويسنده ,
Pages :
6
From page :
129
To page :
134
Abstract :
A polycrystalline Chemical Vapour Deposited (pCVD) diamond film has been processed to implement a bidimensional dosimeter for applications in clinical radiotherapy. The detector, with an area of 2.5×2.5 cm2 and a thickness of 300 μ m , has been equipped with a matrix of electrical contacts (12×12 pixels) on the upper surface. The matrix of pixels has been electrically characterized in dark showing uniformity and very low leakage current. tudy demonstrates that when applied bias is in the range 0–5 V, the pCVD diamond film provides very accurate measurement of dose and dose-rates under clinical 62 MeV proton beam, conventional 6–10 MeV photon beams, and Intensity Modulated Radiotherapy (IMRT) beams.
Keywords :
Polycrystalline CVD diamond dosimeters at low bias , Dose and dose-rate under photon and proton clinical beams , Diamond dosimetry
Journal title :
Astroparticle Physics
Record number :
2010044
Link To Document :
بازگشت