Title of article :
Optimization of the scan protocols for CT-based material extraction in small animal PET/CT studies
Author/Authors :
Yang، نويسنده , , Ching-Ching and Yu، نويسنده , , Jhih-An and Yang، نويسنده , , Bang-Hung and Wu، نويسنده , , Tung-Hsin، نويسنده ,
Pages :
6
From page :
299
To page :
304
Abstract :
We investigated the effects of scan protocols on CT-based material extraction to minimize radiation dose while maintaining sufficient image information in small animal studies. The phantom simulation experiments were performed with the high dose (HD), medium dose (MD) and low dose (LD) protocols at 50, 70 and 80 kVp with varying mA s. The reconstructed CT images were segmented based on Hounsfield unit (HU)-physical density (ρ) calibration curves and the dual-energy CT-based (DECT) method. Compared to the (HU;ρ) method performed on CT images acquired with the 80 kVp HD protocol, a 2-fold improvement in segmentation accuracy and a 7.5-fold reduction in radiation dose were observed when the DECT method was performed on CT images acquired with the 50/80 kVp LD protocol, showing the possibility to reduce radiation dose while achieving high segmentation accuracy.
Keywords :
Dual-energy computed tomography , Material extraction , radiation dose
Journal title :
Astroparticle Physics
Record number :
2010231
Link To Document :
بازگشت