Title of article :
Calibration of a liquid scintillation counter to assess tritium levels in various samples
Author/Authors :
Al-Haddad، نويسنده , , M.N and Fayoumi، نويسنده , , A.H and Abu-Jarad، نويسنده , , F.A، نويسنده ,
Pages :
6
From page :
356
To page :
361
Abstract :
An LKB-Wallac 1217 Liquid Scintillation Counter (LSC) was calibrated with a newly adopted cocktail. The LSC was then used to measure tritium levels in various samples to assess the compliance of tritium levels with the recommended international levels. The counter was calibrated to measure both biological and operational samples for personnel and for an accelerator facility at KFUPM. The biological samples include the bioassay (urine), saliva, and nasal tests. The operational samples of the light ion linear accelerator include target cooling water, organic oil, fomblin oil, and smear samples. Sets of standards, which simulate various samples, were fabricated using traceable certified tritium standards. The efficiency of the counter was obtained for each sample. The typical range of the efficiencies varied from 33% for smear samples down to 1.5% for organic oil samples. A quenching curve for each sample is presented. The minimum detectable activity for each sample was established. Typical tritium levels in biological and operational values are presented. All measured values are far below the recommended international limits.
Keywords :
Oil , Liquid scintillation counter , Calibration , Smear , tritium , bioassay
Journal title :
Astroparticle Physics
Record number :
2010461
Link To Document :
بازگشت