Title of article
Fluence determination by scattering measurements
Author/Authors
Albergo، نويسنده , , S. and Boemi، نويسنده , , D. and Potenza، نويسنده , , R. and Tricomi، نويسنده , , A. and Pillon، نويسنده , , M. and Angarano، نويسنده , , M. and Creanza، نويسنده , , Giovanni D. De Palma، نويسنده , , M.De، نويسنده ,
Pages
7
From page
8
To page
14
Abstract
An alternative method to determine particle fluence is proposed, which is particularly suitable for irradiations with low-energy charged-particle beams. The fluence is obtained by measuring the elastic scattering produced by a composite thin target placed upstream of the sample. The absolute calibration is performed by comparison with the measured radioactivation of vanadium and copper samples. The composite thin target, made of aluminium, carbon and gold, allows not only the fluence to be measured, but also a continuous monitoring of the beam space distribution. Experimental results with a 27 MeV proton beam are reported and compared with Monte Carlo simulations.
Keywords
Fluence silicon radiation damage
Journal title
Astroparticle Physics
Record number
2010488
Link To Document