Title of article :
Comparison of two liquid scintillators used for neutron detection
Author/Authors :
Horvلth، نويسنده , , ء. and Ieki، نويسنده , , K. and Iwata، نويسنده , , Y. and Kruse، نويسنده , , J.J. and Seres، نويسنده , , Z. and Wang، نويسنده , , J. and Weiner، نويسنده , , J. and Zecher، نويسنده , , P.D and Galonsky، نويسنده , , A.، نويسنده ,
Pages :
4
From page :
241
To page :
244
Abstract :
The pulse-shape discrimination capability and the scintillation efficiency of two liquid scintillators, BC519 and NE213 (BC501A) have been compared. Pulse-shape discrimination is not as good with BC519, but for large light pulses it should be adequate in many applications and may be a useful alternative to NE213 because it has a higher ratio of H-to-C- atoms. Within 5% the scintillation efficiencies are the same when the scintillators are excited by protons with energies up to 11 MeV.
Keywords :
Liquid scintillator , Neutron/gamma-ray discrimination