Title of article :
Comparison of LSO samples produced by Czochralsky and modified Musatov methods
Author/Authors :
Antich، نويسنده , , P. and Parkey، نويسنده , , R. and Tsyganov، نويسنده , , E. and Garmash، نويسنده , , V. and Zheleznykh، نويسنده , , I.، نويسنده ,
Pages :
7
From page :
551
To page :
557
Abstract :
This study is based on test results of 30 LSO samples produced by the POLUS Research Institute in Moscow, Russia, for the University of Texas Southwestern Medical Center at Dallas. Samples were produced by the Czochralsky and the modified Musatov methods. Pulse-height spectra from 22Na positron annihilations were analyzed and conclusions are drawn. After some minor corrections to the modified Musatov technology, samples could be recommended for use in PET systems.
Keywords :
Modified Musatov method , inorganic scintillators , LSO crystals
Journal title :
Astroparticle Physics
Record number :
2011297
Link To Document :