Author/Authors :
Schopper، نويسنده , , Florian and Andritschke، نويسنده , , Robert and Shaw، نويسنده , , Henry and Nefzger، نويسنده , , Christian and Zoglauer، نويسنده , , Andreas and Schِnfelder، نويسنده , , Volker and Kanbach، نويسنده , , Gottfried، نويسنده ,
Abstract :
New γ-ray calorimeter have been developed for the MEGA Compton camera. They consist of arrays of small CsI(Tl) scintillator bars read out by Silicon PIN-diodes and low noise, self-triggering frontend electronics. The length of the bars (the thickness of the calorimeter) can be varied for different applications to fit the stopping power needed and the light loss tolerable. In this paper we present calibration results from 2 cm long bars with diodes on one side, and 8 cm long bars with diodes on two opposite sides. Double-sided readout gives 3-D information of interactions which will be used to overcome the limited position resolution in Anger-cameras at high energies. Simpler detection devices like Anger-cameras might finally resolve only the centre of gravity. As events from γ-rays with energies of MeV do extend over several cm, it is a prerequisite for an imaging device to resolve the interaction structure in detail. Combining CsI(Tl) scintillators, Silicon PIN-photodiodes and frontend electronics inside the housing results in a cheap rugged design. While the development in our institute is mainly done for the Compton camera prototype, many other applications are conceivable.