Author/Authors :
Chernenko، نويسنده , , A and Kostenko، نويسنده , , V and Loznikov، نويسنده , , V and Semena، نويسنده , , N and Konev، نويسنده , , S and Rybkin، نويسنده , , B and Paschin، نويسنده , , A and Prokopenko، نويسنده , , I، نويسنده ,
Abstract :
We present current results on the theoretical and experimental studies of optimal cryogenic cooling of gamma-ray spectrometers based on high-purity germanium (HPGe) detectors. We show that the technology of cryogenic heat pipe diodes (HPDs) usually allows one to meet cooling requirements with minimal weight, power consumption and cost. Results of computer modeling and laboratory tests of HPDs, Stirling cooler and complete cooling solutions are presented.