Title of article
Modelling the electro-thermal response of superconducting transition-edge X-ray sensors
Author/Authors
Roth، نويسنده , , P. and Fraser، نويسنده , , G.W. and Holland، نويسنده , , A.D. and Trowell، نويسنده , , S.، نويسنده ,
Pages
13
From page
351
To page
363
Abstract
A two-temperature model for simulating the electrothermal response of superconducting transition-edge sensor (TES) X-ray detectors is presented. The model considers that the absorber and the superconducting layer (SL) of the detector are at different temperatures. The time constants within the absorber and the SL are assumed to be much smaller than all remaining time constants of the TES. The time dependences of the current through the SL, the temperature of the absorber, and the temperature of the SL after the absorption of single photons of different energy are found by numerically solving a set of nonlinear differential equations. In order to verify the model, a calculated current transient is compared with a previously measured pulse. Exploring the parameter space defined by appropriate ratios of electro-thermal time constants, we derive constraints for the asymptotically stable, oscillation-free operation of a TES.
Keywords
superconductor , radiation detector , stability , Electrothermal response
Journal title
Astroparticle Physics
Record number
2011628
Link To Document