Title of article :
Design of a compact structure cancer therapy synchrotron
Author/Authors :
Yang، نويسنده , , J.C. and Shi، نويسنده , , J. and Chai، نويسنده , , W.P. and Xia، نويسنده , , J.W. and Yuan، نويسنده , , Y.J. and Li، نويسنده , , Y.، نويسنده ,
Pages :
4
From page :
19
To page :
22
Abstract :
HIMM, a new compact accelerator facility dedicated to carbon cancer therapy, has been designed and is presently under construction. The synchrotron has a compact structure that exhibits a circumference of only 56.2 m. The charge exchange injection (CEI) method is adopted for synchrotron injection with a carbon-ion energy of 7 MeV/u. The third-order resonance and RF-Knock Out scheme are adopted in this machine. The general design of the machine and injection/extraction simulation results is discussed in this paper.
Keywords :
Charge exchange injection (CEI) , HIMM , Slow extraction
Journal title :
Astroparticle Physics
Record number :
2012296
Link To Document :