Title of article
Influence of the active area size and read-out method on the timing performance of SiPMs coupled to LYSO scintillators
Author/Authors
Tarolli، نويسنده , , Alessandro and Gola، نويسنده , , Alberto and Ferri، نويسنده , , Alessandro and Pro، نويسنده , , Tiziana and Serra، نويسنده , , Nicola and Zorzi، نويسنده , , Nicola and Piemonte، نويسنده , , Claudio، نويسنده ,
Pages
3
From page
77
To page
79
Abstract
In this work we report on the coincidence resolving time performance of SiPMs with different sizes, produced at FBK, coupled to the same LYSO scintillators. The measurements are performed both with and without the differential leading edge discriminator at three different temperatures, 20 °C, 0 °C and −20 °C. The photo-detectors feature an active area of 2×2 mm2 and 4×4 mm2. The scintillators have a cross-section of 1.8×1.8 mm2 and height of 10 mm. The measurements show that, once we eliminate the effect of noise on the timing measurements, we obtain similar coincidence resolving times for the two SiPM sizes considered. This means that the SiPM capacitance, at least up to 4×4 mm2, is not a limiting factor.
Keywords
SiPM , CRT , DLED
Journal title
Astroparticle Physics
Record number
2012614
Link To Document