Title of article :
Response of monitoring instruments to high-energy photon radiation
Author/Authors :
Haridas، نويسنده , , G and Thakker، نويسنده , , K.K and Pradhan، نويسنده , , S.D. and Nayak، نويسنده , , A.R and Bhagwat، نويسنده , , A.M، نويسنده ,
Pages :
5
From page :
624
To page :
628
Abstract :
Response of commercially available monitoring instruments to high-energy photon radiation was studied under the stored beam condition of a few milliamperes in the storage ring of the Synchrotron Radiation Source, INDUS-I, at Centre for Advanced Technology (CAT), Indore. The storage ring has a circumference of 18.96 m, where electrons at 450 MeV are stored for a few hours, during which the emitted synchrotron radiation is exploited for scientific research and other applications. Radiation environment near storage ring has bremsstrahlung photons of various energies (maximum 450 MeV). A study has indicated underestimation of dose by conventional radiation monitoring instruments by a factor of 2–4. Response after transmission of photons through massive shield was also studied, which indicated spectral degradation and good response by the survey meters.
Keywords :
Response , High-energy electron accelerator , Instruments , Bremsstrahlung , buildup , Depth dose
Journal title :
Astroparticle Physics
Record number :
2012616
Link To Document :
بازگشت