Title of article
Gain calibration of the ALICE TRD using the decay of 83mKr by internalconversion
Author/Authors
Stiller، نويسنده , , Johannes، نويسنده ,
Pages
3
From page
20
To page
22
Abstract
For an early calibration of the ALICE TRD on the level of individual readout pads a dedicated calibration run with a Krypton source was carried out. We recorded 2.28×109 decays of metastable 83mKr in the 6.46×105 readout-pads of the 10 TRD super modules presently installed in the ALICE experiment [1] at CERN. In our analysis the gain uniformity within each chamber as well as the energy resolution of all installed chambers were obtained. This is crucial for good online particle identification and thus triggering. In an iterative analysis step, these gain factors converged to 1% gain uniformity at the readout pad level. Additionally, a comparison between the obtained results and corresponding findings from measurements during chamber construction was performed and good agreement was found. The pad gain factors are uploaded into the front-end electronics for online gain correction to ensure stable trigger conditions.
Keywords
GAIN , Calibration , krypton , Transition radiation detector , Drift chamber , ALICE
Journal title
Astroparticle Physics
Record number
2012826