Author/Authors :
Pawe?czak، نويسنده , , I.A. and Ouedraogo، نويسنده , , S.A. and Glenn، نويسنده , , A.M. and Wurtz، نويسنده , , R.E. and Nakae، نويسنده , , L.F.، نويسنده ,
Abstract :
Pulse shape discrimination capability based on the charge integration has been investigated for liquid scintillator EJ-309. The effectiveness of neutron– γ discrimination in 4-in. diameter and 3-in. thick EJ-309 cells coupled with 3-in. photomultiplier tubes has been carefully studied in the laboratory environment and compared to the commonly used EJ-301 liquid scintillator formulation. Influences of distortions in pulse shape caused by 13.7-m long cables necessary for some remote operations have been examined. The parameter space for an effective neutron– γ discrimination for these assays, such as position and width of a gate used for integration of the delayed light, has been explored.
Keywords :
Liquid scintillator , EJ-309 , Pulse shape discrimination , Fast-neutron detection