Title of article :
Methods for precise photoelectron counting with photomultipliers
Author/Authors :
Dossi، نويسنده , , R. and Ianni، نويسنده , , A. and Ranucci، نويسنده , , G. and Smirnov، نويسنده , , O.Ju.، نويسنده ,
Pages :
15
From page :
623
To page :
637
Abstract :
A series of measurements has been performed on a THORN EMI 9351 phototube in order to investigate its response to a low light intensity. Precise procedures to determine the intensity of the incident photon flux have been developed and compared. The data show that the various approaches give consistent and reliable results, thus allowing the precise calibration of the device for applications of photon counting.
Keywords :
Photomultiplier , scintillator , Calibration , Photoelectron
Journal title :
Astroparticle Physics
Record number :
2012967
Link To Document :
بازگشت