• Title of article

    Transconductance calibration of n-XYTER 1.0 readout ASIC

  • Author/Authors

    Sorokin، نويسنده , , I. S. Balog ، نويسنده , , Peter T. and Krylov، نويسنده , , V. and Schmidt، نويسنده , , C.J.، نويسنده ,

  • Pages
    5
  • From page
    136
  • To page
    140
  • Abstract
    After an overview of n-XYTER 1.0—a high-gain, charge-sensitive, self-triggering readout ASIC—the experimental work on its transconductance calibration is presented in details. nce charge pulses were generated by applying a voltage step over a capacitor. Special care was taken to achieve the best possible precision and to check for potential systematic errors. The calibration was done on 3 chips and many channels. Average calibration curves were determined for both positive and negative polarities and the uncertainties were estimated. y, correctness of the calibration was confirmed by a measurement with an alternative signal source—a silicon detector, exposed to γ - radiation of 241Am.
  • Keywords
    Transconductance calibration , Charge-sensitive amplifier calibration , Charge pulse generator , Self-triggered , n-XYTER
  • Journal title
    Astroparticle Physics
  • Record number

    2013107