Title of article
Event identification in 3He proportional counters using risetime discrimination
Author/Authors
Langford، نويسنده , , T.J. and Bass، نويسنده , , C.D. and Beise، نويسنده , , E.J. and Breuer، نويسنده , , H. and Erwin، نويسنده , , D.K. and Heimbach، نويسنده , , C.R. and Nico، نويسنده , , J.S.، نويسنده ,
Pages
7
From page
51
To page
57
Abstract
We present a straightforward method for particle identification and background rejection in 3He proportional counters for use in neutron detection. By measuring the risetime and pulse height of the preamplifier signals, one may define a region in the risetime versus pulse height space where the events are predominately from neutron interactions. For six proportional counters surveyed in a low-background environment, we demonstrate the ability to reject alpha-particle events with an efficiency of 99%. By applying the same method, we also show an effective rejection of microdischarge noise events that, when passed through a shaping amplifier, are indistinguishable from physical events in the counters. The primary application of this method is in measurements where the signal-to-background for counting neutrons is very low, such as in underground laboratories.
Keywords
Neutron detectors , Thermal neutrons , Pulse shape discrimination , Particle identification , Microdischarges , Helium proportional counters
Journal title
Astroparticle Physics
Record number
2013162
Link To Document