Title of article
A sensitive high resolution neutron imaging detector
Author/Authors
Cheng، نويسنده , , Y.T and Mildner، نويسنده , , D.F.R، نويسنده ,
Pages
8
From page
452
To page
459
Abstract
We present a two-step method for a high-resolution neutron area detector in which a metal foil with a high absorption cross section acts as a neutron to charged particle converter. After the foil is activated, the neutron field information is transferred from the converter foil to an imaging device by autoradiography. This two-step process using a storage photostimulable phosphor avoids interferences from non-neutron radiation found in the radiation field.
Keywords
Imaging plate detector , Neutron beam profiles , Photostimulable luminescence , Neutron field distribution , Portable neutron area detector , Storage photostimulable phosphor , Background interference , Converter foil activation
Journal title
Astroparticle Physics
Record number
2013410
Link To Document