• Title of article

    Electron-beam diagnosis with Youngʹs interferometer in soft X-ray region

  • Author/Authors

    Takayama، نويسنده , , Yasuhiro and Shiozawa، نويسنده , , Hidetsugu and Takaya، نويسنده , , Noriko and Miyahara، نويسنده , , Tsuneaki and Tai، نويسنده , , RenZhong، نويسنده ,

  • Pages
    5
  • From page
    217
  • To page
    221
  • Abstract
    We report an emittance measurement with undulator radiation performed at BL-16B of the Photon Factory, KEK. The spatial coherence (visibility) and the beam size of undulator radiation are measured with a Youngʹs interferometer and a wire scanner in the soft X-ray region, of about 40–200 eV. The horizontal emittance is estimated as around 40 nm rad, which is consistent with the design value 36 nm rad. The vertical emittance is estimated as larger than 1.5 nm rad. This value is too large if the coupling is assumed to be 2%. The discrepancy in the vertical direction is due to an incomplete adjustment of the direction of a double slit and phase randomization by a dirty grating in the Youngʹs interferometer.
  • Keywords
    Undulator , spatial coherence , Youngיs interferometer
  • Journal title
    Astroparticle Physics
  • Record number

    2013504