• Title of article

    Measurement of the total-to-peak ratio of a low-energy germanium gamma-ray detector

  • Author/Authors

    Korun، نويسنده , , M.، نويسنده ,

  • Pages
    8
  • From page
    245
  • To page
    252
  • Abstract
    The spatial dependence of the total-to-peak ratio for a closed-end coaxial n-type detector with a beryllium window was measured. For the measurements point sources containing pure X-ray emitters and X- and gamma-ray emitters, emitting photons in the energy range between 5 and 1115 keV were used. Departures of up to 10% from a total-to-peak ratio independent of the source position were found. They are attributed to the escape of germanium X-rays and scattered photons from the sensitive volume of the detector and to detection of radiation scattered in the insensitive parts of the detector.
  • Keywords
    X-rays , gamma rays , Low-energy gamma ray spectrometer , Total-to-peak ratio , Co-incidence summary , X-ray Absorption
  • Journal title
    Astroparticle Physics
  • Record number

    2013753