• Title of article

    Gas proportional scintillation counters with a CsI-covered microstrip plate UV photosensor for high-resolution X-ray spectrometry

  • Author/Authors

    Veloso، نويسنده , , J.F.C.A. and dos Santos، نويسنده , , J.M.F and Conde، نويسنده , , C.A.N، نويسنده ,

  • Pages
    9
  • From page
    253
  • To page
    261
  • Abstract
    A xenon-filled gas proportional scintillation counter for X-ray spectrometry is described. The detector uses a microstrip plate covered with a thin CsI film as a photosensor rather than the usual photomultiplier tube. The achieved energy resolution for 5.9 keV X-rays is 12% when the microstrip is operated within the xenon volume and 10.5% when the microstrip is operated in a P-10 environment, separated from the xenon volume by a quartz window.
  • Keywords
    Gas proportional scintillation counter , MSGC , CsI photocathode , X-rays
  • Journal title
    Astroparticle Physics
  • Record number

    2013754