• Title of article

    The spatial response of CdZnTe gamma-ray detectors as measured by gamma-ray mapping

  • Author/Authors

    Brunett، نويسنده , , B.A. and Van Scyoc، نويسنده , , J.M. and Schlesinger، نويسنده , , T.E and James، نويسنده , , R.B.، نويسنده ,

  • Pages
    9
  • From page
    76
  • To page
    84
  • Abstract
    We have developed a system to measure the spatial response of cadmium zinc telluride (CZT) radiation detectors. Using this system we have measured the response of several novel detector designs including several variations of the unipolar design. We have observed a wide range of energy resolution and efficiency among the different device designs. Each design has unique strengths and weaknesses which affect the device performance. In addition to design effects on performance, several instances of poor material uniformity degrading the device performance have been observed. In this paper we will discuss the spatial detector response focusing on the effects of the detector design. Where appropriate, we will also discuss the observed effects of material uniformity on device performance.
  • Keywords
    cadmium zinc telluride , Mapping , Gamma-ray detectors , Unipolar , Spectrometer
  • Journal title
    Astroparticle Physics
  • Record number

    2013784