Title of article :
Spatial resolution in Micromegas detectors
Author/Authors :
Derré، نويسنده , , J. and Giomataris، نويسنده , , Y. and Zaccone، نويسنده , , H. and Bay، نويسنده , , A. and Perroud، نويسنده , , J.-P. and Ronga، نويسنده , , F.، نويسنده ,
Pages :
9
From page :
523
To page :
531
Abstract :
The performance of a telescope of Micromegas detectors has been studied in a pion beam at the CERN PS. With a gas filling of CF4 and 20% isobutane and with a strip pitch of 100 μm an accuracy of 14±3 μm on the spatial resolution has been measured at normal incidence. A simulation demonstrates that the resolution is limited by the size of the holes of the mesh of the detector and could be reduced to 11 μm in the same conditions with smaller holes. Even further improvement down to 8.5 μm is feasible for the same gas with an optimized 75 μm strip pitch.
Keywords :
Charged particle tracking in gaseous detector , Spatial accuracy , Space resolution , Space accuracy , Spatial resolution , Micromegas detector
Journal title :
Astroparticle Physics
Record number :
2014015
Link To Document :