Title of article :
Dose-dependent X-ray measurements using a 64×64 hybrid GaAs pixel detector with photon counting
Author/Authors :
Schwarz، نويسنده , , C. and Campbell، نويسنده , , M. and Goeppert، نويسنده , , R. and Ludwig، نويسنده , , J. and Mikulec، نويسنده , , B. and Rogalla، نويسنده , , M. and Runge، نويسنده , , K. and Soeldner-Rembold، نويسنده , , A. and Smith، نويسنده , , K.M. and Snoeys، نويسنده , , W. E. Watt، نويسنده , , J.، نويسنده ,
Pages :
6
From page :
91
To page :
96
Abstract :
New developments in medical imaging head towards semiconductor detectors flip-chip bonded to CMOS readout chips. In this work, detectors fabricated on SI-GaAs bulk material were bonded to Photon Counting Chips.11Developed in the framework of the MEDIPIX collaboration: CERN, Universities of Freiburg, Glasgow, Naples and Pisa. PCC consists of a matrix of 64×64 identical square pixels (170 μm×170 μm) with a 15-bit counter in each cell. We investigated the imaging properties of these detector systems under exposure of a dental X-ray tube. First, a dose calibration of the X-ray tube was performed. Fixed pattern noise in flood exposure images was determined for a fixed dose and an image correction method, which uses a gain map, was applied. For characterising the imaging properties, the signal-to-noise ratio (SNR) was calculated as function of exposure dose. Finally, the dynamic range of the system was estimated.
Keywords :
GaAs , MEDICAL IMAGING , Pixel detector , X-rays , Photon counting chip , Medipix
Journal title :
Astroparticle Physics
Record number :
2014045
Link To Document :
بازگشت