Author/Authors :
Peng، نويسنده , , Song and Liu، نويسنده , , Zhiguo and Sun، نويسنده , , Tianxi and Ma، نويسنده , , Yongzhong and Sun، نويسنده , , Weiyuan and Zhao، نويسنده , , Weigang and He، نويسنده , , Jialin and Zhao، نويسنده , , Guangcui and Ding، نويسنده , , Xunliang، نويسنده ,
Abstract :
Adjustment of the confocal configuration of capillary X-ray optics using a solution of CuSO4 as a liquid secondary target is presented. Compared with the theoretical value of volume size, the relative error of the adjustment with a liquid secondary target was 9.8%, and this value was more accurate than that observed with a metal secondary target. The precision of the adjustment using a liquid secondary target was better than 7%. The adjustment process for confocal configuration with a liquid secondary target was more convenient and timesaving because there was no need for an extra electric stage to adjust the liquid secondary target.
Keywords :
Confocal configuration , Capillary X-ray optics , Polycapillary X-ray optics , Liquid secondary target