• Title of article

    Study of a new scatter rejection technique in digital radiography

  • Author/Authors

    Bottari، نويسنده , , S. and Ciocci، نويسنده , , M.A. and Fortunato، نويسنده , , M. Carmen Maestro، نويسنده , , P. and Malakhov، نويسنده , , N. and Marrocchesi، نويسنده , , P.S. and Meucci، نويسنده , , M. and Millucci، نويسنده , , V. and Paoletti، نويسنده , , R. and Scribano، نويسنده , , A. and Turini، نويسنده , , N.، نويسنده ,

  • Pages
    3
  • From page
    410
  • To page
    412
  • Abstract
    A new technique for digital mammography based on the use of a collimator and an anti-scatter grid coupled with a mosaic detector has been studied with a Monte Carlo program. The simulation, with a low-energy spectrum X-ray beam and a breast phantom, provides a quantitative assessment of the capability of the method to reduce the physical background of the image due to scattering in the body, without introducing image artifacts. With minor modifications to the existing X-ray facilities, the method could also be applied to area detectors. To verify the results of the simulation, an experimental setup based on a CCD camera coupled via a fiber optic plate to a CsI(Tl) scintillator is under test.
  • Keywords
    digital mammography , digital radiography
  • Journal title
    Astroparticle Physics
  • Record number

    2014518