Title of article :
Neutron self-shielding with k0-NAA irradiations
Author/Authors :
Chilian، نويسنده , , C. and Chambon، نويسنده , , R. and Kennedy، نويسنده , , G.، نويسنده ,
Pages :
4
From page :
429
To page :
432
Abstract :
A sample of SMELS Type II reference material was mixed with powdered Cd-nitrate neutron absorber and analysed by k0 NAA for 10 elements. The thermal neutron self-shielding effect was found to be 34.8%. When flux monitors were irradiated sufficiently far from the absorbing sample, it was found that the self-shielding could be corrected accurately using an analytical formula and an iterative calculation. When the flux monitors were irradiated 2 mm from the absorbing sample, the calculations over-corrected the concentrations by as much as 30%. It is recommended to irradiate flux monitors at least 14 mm from a 10 mm diameter absorbing sample.
Keywords :
NEUTRON ACTIVATION ANALYSIS , K0 method , Neutron self-shielding
Journal title :
Astroparticle Physics
Record number :
2014801
Link To Document :
بازگشت